Abstract
Background
Accurate malaria diagnosis is key for patient management, surveillance, and control. Automated microscopy can overcome variation observed among microscopists and is a promising new tool for diagnosis. The Noul miLab integrates smear preparation, staining, imaging, and AI-supported parasite detection in a portable device.
Methods
2201 samples were collected from febrile patients across two sites in Ethiopia, where
P. falciparum
and
P. vivax
are frequent, and Ghana, where
P. falciparum
transmission is intense. Samples were screened by local microscopy at the health center, miLab, and rapid diagnostic test. qPCR and expert microscopy were used as gold standard.
Results
Across the three sites in Ehtiopia and Ghana, miLab reached a sensitivity for
P. falciparum
diagnosis of 96.3% (335/348) using expert microscopy as gold standard, and of 97.4% (298/306) using qPCR-positive infections at densities >200 parasites/µL as gold standard. Across two sites in Ethiopia, the sensitivity of miLab for
P. vivax
was 96.8% (399/412) using expert microscopy as gold standard, and 95.9% (419/437) using qPCR-positive infections at densities >200 parasites/µL as gold standard. Specificity compared to qPCR was 98.8% (1057/1070) for
P. falciparum
and 97.8% (617/631) for
P. vivax
. The miLab was significantly more sensitive than microscopy conducted at the health center. In Ethiopia, the miLab assigned the correct species to 99.3% (147/148)
P. falciparum
and 96.5% (304/315)
P. vivax
mono-infections infections where a species was determined.
Conclusions
The miLab automated microscope shows high sensitivity and specificity for
P. falciparum
and
P. vivax
diagnosis.