Evaluation of a continuous neonatal temperature monitor for low-resource settings: a device feasibility pilot study

Domain:

healthcare
Creator:
SonMarMegZ.
Publisher:
BMJ
Host:
Objective Evaluate a novel continuous temperature monitor in a low-resource neonatal ward. Design We developed a low-cost continuous neonatal temperature monitor (NTM) for use in low-resource settings. Accuracy of NTM was initially assessed in the laboratory. Clinical evaluation then was performed in a neonatal ward in a central hospital in Malawi; eligible neonates (<1 week of age) were recruited for continuous temperature monitoring with NTM and a Philips Intellivue MP30 Patient Monitor. Interventions and outcome measures The temperature probes of NTM and the reference patient monitor were attached to the infant’s abdomen, and core temperature was continuously recorded for up to 3 hours. Axillary temperatures were taken every hour. We compared temperatures measured using NTM, the patient monitor and the axillary thermometer. Results Laboratory temperature measurements obtained with NTM were within 0.059°C (range: −0.035°C to 0.195°C) of a reference thermometer. A total of 39 patients were recruited to participate in the clinical evaluation of NTM; data from four patients were excluded due to faulty hardware connections. The mean difference in measured temperatures between the NTM and the Intellivue MP30 was −0.04°C (95% CI −0.52°C to 0.44°C). Conclusion NTM meets ISO 80601-2-56 standards for accuracy and is an appropriate, low-cost continuous temperature monitor for neonatal wards in low-resource settings. Trial registration numbers NCT03965312 and NCT03866122 .

Visit

doi.org

Licenses

http://creativecommons.org/licenses/by-nc/4.0/

Similar

Acceptability and Feasibility of a Low-Cost Device for Gestational Age Assessment in a Low-Resource Setting: Qualitative StudyEvaluation of ‘TRY’: an algorithm for neonatal continuous positive airways pressure in low-income settingsA New Solar-Powered Blood Pressure Measuring Device for Low-Resource SettingsReusable Core Needle Biopsy Device for Low-Resource SettingsDevelopment of a novel device for objective respiratory rate measurement in low-resource settingssj-docx-1-dhj-10.1177_20552076221109060 - Supplemental material for A novel low-cost bubble continuous positive airway pressure device with pressure monitoring and controlling system for low resource settings

Acceptability and Feasibility of a Low-Cost Device for Gestational Age Assessment in a Low-Resource Setting: Qualitative Study

Background Ultrasound for gestational age (GA) assessment is not routinely available i

Evaluation of ‘TRY’: an algorithm for neonatal continuous positive airways pressure in low-income settings

Background Non-invasive respiratory support using bubble continuous positive airway pressure (bCPAP

A New Solar-Powered Blood Pressure Measuring Device for Low-Resource Settings

The management of high blood pressure (BP) is particularly inadequate in low-income countries, where

Reusable Core Needle Biopsy Device for Low-Resource Settings

Purpose The core needle biopsy (CNB) procedure is an essential part of breast cancer diagnosis. Cu

Development of a novel device for objective respiratory rate measurement in low-resource settings

Objective To evaluate a novel device (Respimometer) for objective measurement of

sj-docx-1-dhj-10.1177_20552076221109060 - Supplemental material for A novel low-cost bubble continuous positive airway pressure device with pressure monitoring and controlling system for low resource settings

Supplemental material, sj-docx-1-dhj-10.1177_20552076221109060 for A novel low-cost bubble continuou